HPLC–MS/MS analysis of peramivir in rat plasma: Elimination of matrix effect using the phospholipid‐removal solid‐phase extraction method

Skip to Navigation

EarlyView Article

  • Published: Oct 25, 2017
  • Author: Mingdao Lei, Wei Gan, Yongbing Sun
  • Journal: Biomedical Chromatography

Abstract

A simple HPLC–MS/MS method has been developed for the determination of peramivir in rat plasma in the present study. The analytes were separated on a C18 column (50 × 2.1 mm, 1.7 μm) and a triple‐quadrupole mass spectrometer equipped with an electrospray ionization source was applied for the detection. A phospholipid‐free cartridge solid‐phase extraction was used to pretreat the plasma and eliminate the endogenous phospholipid. The in‐source collision‐induced dissociation approach showed that this pretreatment could result in negligible ion suppression from the extracted sample and could produce cleaner samples when compared with the protein precipitation. The method was linear over the concentration range of 0.12–1200.0 ng/mL for peramivir. The method was validated and successfully applied to a pharmacokinetic study after peramivir was orally and intravenously administered to Sprague–Dawley rats.